In reviewSelect replays by recording and flag exports that lack a recording referenceCHG-1106
Current state
The name-based lookup returns the first recording for the device. Selecting B can replay A because their headers name the same recorder.
Proposed change
The alternative resolves the selected manifest entry to the original recording before replay. The B selection now returns B’s distinct sequence.
Reason
Matching a device name can produce a plausible replay of the wrong recording. The separate bench comparison proves the new lookup for the two named entries; it supplies no missing reference for the older handoff.
Decision owner
Recorder software reviewer
Last activity
Nov 22, 2026, 4:45 PM UTC
Applies to
Weka data recorder · Rev A · WR-101
Decision
Choose what replay should show for an older export with no capture reference. Do not silently fall back to the first recording.
Follow-up
Unavailable
Notes
In reviewAdd the original recording reference to the WR-101 replay exportCHG-832
Current state
The handoff copy names the recorder, WR-101. It gives the reviewer no way to look up the recording that was replayed.
Proposed change
Add the reference from the capture manifest to the handoff copy. Keep the originally filed export.
Reason
A recorder can produce more than one recording. Its serial cannot identify this recording on its own.
Decision owner
Recorder integration reviewer
Last activity
Unavailable
Applies to
Weka data recorder · Rev A · WR-101
Decision
Check the proposed reference against the capture manifest before approving the WR-101 handoff copy.